Agilent Technologies 856290216 service manual Ramp Counter

Models: 856290216

1 482
Download 482 pages 3.94 Kb
Page 301
Image 301

comparator U512 does not toggle back and forth during a conversion, suspect either U512 or one of the clipping diodes (CR500/CR501).

Note

Because currents are being summed at U512 pins 2 and 3, voltage levels at

 

these points are difficult to interpret.

 

 

Ramp Counter

Refer to function block D of A3 Interface Assembly Schematic Diagram in the

HP 8560 E-Series Spectrum Analyzer Component Level Information.

The ramp counter is used for sweeps with widths greater than 2.0 MHz times N. The analog sweep ramp is compared to the digital ramp counter. When the analog sweep ramp exceeds the DAC output generated for that ramp counter setting, HRAMP-COMP toggles high, indicating the end of a bucket. The ramp counter counts horizontal buckets. There are 601 buckets per sweep, so the ramp (bucket) counter counts from 0 to 600. The ramp counter is incremented by HRST-PK-ENA.

1. Press @GZi] on the spectrum analyzer and set the controls as follows:

Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MHz Detector mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SAMPLE

2.For spans greater than 2.0 MHz times NO, HODD (A3U525 pin 3) is a square wave with a period defined by (2 x sweep time/600). For example, for a 6 s sweep time, HODD has a period of 20 ms. The ramp (bucket) counter will be odd every other bucket.

ADC/lnterface Section 6-27

Page 301
Image 301
Agilent Technologies 856290216 service manual Ramp Counter