502

AMPL

.3M

Unable to adjust amplitude of 300 kHz resolution bandwidth.

503

AMPL

1M

Unable to adjust amplitude of 1 MHz resolution bandwidth.

504

AMPL

30K

Unable to adjust amplitude of 30 kHz resolution bandwidth.

505

AMPL

.lM

Unable to adjust amplitude of 100 kHz resolution bandwidth.

506

AMPL

.3M

Unable to adjust amplitude of 300 kHz resolution bandwidth.

507 AMPL

1M

Unable to adjust amplitude of 1 MHz resolution bandwidth.

508

AMPL

30K

Unable to adjust amplitude of 30 kHz resolution bandwidth.

 

 

 

Insufficient gain during LC bandwidth calibration.

509

AMPL .lM

Unable to adjust amplitude of 100 kHz resolution bandwidth.

 

 

 

Insufficient gain during LC bandwidth calibration.

510 AMPL .3M

Unable to adjust amplitude of 300 kHz resolution bandwidth.

 

 

 

Insufficient gain during LC bandwidth calibration.

511 AMPL 1M

Unable to adjust amplitude of 1 MHz resolution bandwidth.

 

 

 

Insufficient gain during LC bandwidth calibration.

512 RBW <300

Unable to adjust resolution bandwidths less than 300 Hz. Insufficient

 

 

 

gain during crystal bandwidth calibration.

513

RBW

300

Unable to adjust 300 Hz resolution bandwidth. Insufficient gain during

 

 

 

crystal bandwidth calibration.

514 RBW IK

515 RBW 3K

516 RBW 10K

Unable to adjust 1 kHz resolution bandwidth. Insufficient gain during crystal bandwidth calibration.

Unable to adjust 3 kHz resolution bandwidth. Insufficient gain during crystal bandwidth calibration.

Unable to adjust 10 kHz resolution bandwidth. Insufficient gain during crystal bandwidth calibration.

517 RBW 100 Unable to adjust 100 Hz resolution bandwidth. Crystal sweep problem.

518 RBW 300

519 RBW IK

520 RBW 3K

521 RBW 1OK

522 RBW 1OK

Unable to adjust 300 Hz resolution bandwidth. Crystal sweep problem.

Unable to adjust 1 kHz resolution bandwidth. Crystal sweep problem.

Unable to adjust 3 kHz resolution bandwidth. Crystal sweep problem.

Unable to adjust 10 kHz resolution bandwidth. Crystal sweep problem.

Unable to adjust symmetry of 10 kHz resolution bandwidth in first crystal pole.

523 RBW 10K

524 RBW 10K

525 RBW 10K

Unable to adjust symmetry of 10 kHz resolution bandwidth in second crystal pole.

Unable to adjust symmetry of 10 kHz resolution bandwidth in third crystal pole.

Unable to adjust symmetry of 10 kHz resolution bandwidth in fourth crystal pole.

General Troubleshooting 7-33

Page 252
Image 252
Agilent Technologies 856290216 service manual Ampl 1M