13.Figure 9-17 illustrates normal operation. Severe failures (slope error greater than 30 percent) and subtle 3 kHz resolution bandwidth errors (less than 30 percent) indicate a problem with A4U802, A4U803, A4U804, or A4U106.

14.Severe failure of the bandwidth accompanied by subtle errors in the output signal indicate an A5 failure.

15.Set the HP 8566A/B controls as follows:

Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.710 MHz Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz Videobandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1kHz Sweep time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200ms

16.On the HP 85623 spectrum analyzer, press FULL IF AI)3 . When the message IF ADJUST STATUS: 1 kHz RBW appears, press (j-j on the HP 8566A/B.

17.Figure 9-18 illustrates normal operation. Severe failures (slope error greater than 30 percent) and subtle 3 kHz resolution bandwidth errors (less than 30 percent) indicate a problem with A4U802, U803, U804, or U106.

18.On the HP 85623 spectrum analyzer, press FULL IF RDJ . When the message IF ADJUST STATUS: 300 Hz RBW appears, press (s) on the HP 8566A/B.

19.Figure 9-19 illustrates normal operation. Severe failures (slope error >30 percent) and 3 kHz resolution bandwidth errors (less than 30 percent) indicate a problem with A4U802, U803, U804, or U106.

20.Reconnect W52 (white) to A4J8.

O u t p u t Wavrform. LB kHZ ReaOlutiOn Bandwidth

REF - 4 0 . 0 clBm ATTEN i0 08

rp

i de/

I

I

I

I

I

1,

I

I

I

I

I

C E N T E R i 0 . 8 0 0 0 0 0 Ml-It

 

 

 

 

 

 

S P A N 0 HZ

 

RES BW i00 kl-lz

 

VBW

10 kHz

 

 

SWP

5 0 . 0

mast

Figure 9-16. Output Waveform, 10 kHz Resolution Bandwidth

IF Section 9-35

Page 350
Image 350
Agilent Technologies 856290216 service manual Output Waveform, 10 kHz Resolution Bandwidth If Section