Table 11-2. Automatic Fault Isolation References (continued)

 

Suspected Circuit Indicated

Manual Procedure to Perform

 

by Automatic Fault Isolation

 

Check

offset

span accuracy

First LO Span Problems 52 MHz (step 8)

Check

phase/frequency detector

Unlocked Reference PLL (steps 17-22)

Check path to phase/frequency detector

Unlocked Offset PLL (steps 3-7,14-f9)

Check sampler drive output of A7

Unlocked YTO PLL (steps 18-21)

Check

sampler IF

Unlocked YTO PLL (steps 22-26)

Check

sampler/sampler IF operation

Sampler and Sampler IF (steps l-15)

Check

span

attenuator

First LO Span Problems (All Spans) (steps 6-13)

Check

sweep

generator

Sweep Generator Circuit

Check sweep + tune multiplier

YTF Driver Circuit (steps 4-9)

Check the 600 MHz reference loop amplifier

Unlocked Reference PLL (steps 23-26)

Check the YTO loop

Unlocked YTO PLL

Check YTO FM coil driver

First LO Span Problems (2.01 MHz to 20 MHz)

 

 

 

(Step 6)

Check YTO FM coil driver and main loop

Unlocked YTO PLL (steps 3540)

error voltage

driver

 

 

 

 

 

Synthesizer Section 1l-5

Page 379
Image 379
Agilent Technologies 856290216 service manual Synthesizer l-5