Table 1-4. Recommended Test Equipment (continued)

Instrument

Accessories

Directional bridge

Directional coupler

(two required)

10 dB step attenuator

1 dB step attenuator

20 dB fixed attenuator

10 dB fixed attenuator

Reference attenuator

Critical Specifications for Equipment Substitution

Frequency range: 1 to 80 MHz

Coupling: 6 dB (nominal)

Maximum coupling deviation: < 1 dB (nominal)

Directivity: 40 dB minimum

Impedance: 50 R (nominal)

Frequency range: 2.0 to 8.1 GHz

Coupling: 16.0 dB (nominal)

Maximum coupling deviation: fl dB (nominal)

Directivity: 14 dB minimum

Flatness: 0.75 dB maximum

VSWR: <1.45

fnsertion loss: <1.3 dB

Attenuation range: 30 dB

Frequency range: dc to 80 MHz

Connectors: BNC(f)

Attenuation range: 12 dB

Frequency range: dc to 80 MHz

Connectors: BNC(f)

Frequency range: dc to 13.2 GHz

Attenuation accuracy: <fl dB

Maximum SWR: 1.2 (dc to 2.9 GHz)

Frequency range: dc to 13.2 GHz

Attenuation accuracy: <f0.6 dB

Maximum SWR: 1.2 (dc to 2.9 GHz)

Supplied with HP 8481D

Recommended Use Model

HP 8721A

P

0955-0098

P

HP 355D

p,v

HP 355C

p,v

HP 8491B p,v Option 020

HP 8491B p,v Option 010

HP 11708A

P,A

* Part of microwave workstation

P = performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting; V = operation verification

General Information l-1 3

Page 36
Image 36
Agilent Technologies 856290216 service manual Instrument Accessories