System Errors (750 to 759)

 

 

These errors often require troubleshooting the A2 controller and A3

 

 

interface assemblies.

750

SYSTEM

Hardware/firmware interaction, zero divide. Check for other errors.

751

SYSTEM

Hardware/firmware interaction, floating point overflow. Check for

 

 

other errors.

752

SYSTEM

Hardware/firmware interaction, floating point underflow. Check for

 

 

other errors.

753

SYSTEM

Hardware/firmware interaction, log error. Check for other errors.

754

SYSTEM

Hardware/firmware interaction, integer overflow. Check for other

 

 

errors.

755

SYSTEM

Hardware/firmware interaction, square root error. Check for other

 

 

errors.

756 SYSTEM

Hardware/firmware interaction, triple overflow. Check for other errors.

757 SYSTEM

Hardware/firmware interaction, BCD overflow. Check for other errors.

758

SYSTEM

Unknown system error.

759

SYSTEM

Hardware/firmware interaction. Code invoked for wrong instrument.

Fast ADC Error (760)

This error applies only to spectrum analyzers with fast ADC (Option 007).

760 NO FADC The FADC board did not respond properly to initialization commands.

Option Module Errors (800 to 899)

These error codes are reserved for option modules, such as the HP 85629 test and adjustment module and the HP 85620A mass memory module. Refer to the option module manual for a listing of error messages.

7-40 General Troubleshooting

Page 259
Image 259
Agilent Technologies 856290216 service manual System Errors 750 to, Fast ADC Error, Option Module Errors 800 to