Procedure 2. Al Front Frame/Al8 CRT

A18Wl

A18Wl

1

n

sm627e

Figure 4-5. Installing the CRT and front frame Assemblies

Replacement

Note

Use care when handling the glass CRT EM1 shield. The glass may be cleaned

 

using thin-film cleaner (HP part number 8500-2163) and a lint-free cloth.

 

When installing the glass shield, face the side of the glass with the silver

 

coated edge towards the inside of the spectrum analyzer.

 

 

1.Place the spectrum analyzer on its right side frame with the front end extending slightly over the front of the work bench.

2.Gently place the Al8 CRT assembly into the Al front frame assembly as illustrated in Figure 4-6.

3.Place the front frame and CRT assemblies into the spectrum analyzer, using caution to avoid pinching any cables.

4.Dress the A18Wl trace-align wires between the CRT assembly mounts and the A6 power supply top shield.

4-10 Assembly Replacement

Page 135
Image 135
Agilent Technologies 856290216 service manual Replacement, Installing the CRT and front frame Assemblies