11. 10 MHz Reference Adjustment-OCXO (Non-Option 103)

Assembly Adjusted

A21 OCXO assembly

Note

Replacement oscillators are factory adjusted after a complete warmup and

 

after the specified aging rate has been achieved. Thus, readjustment should

 

typically not be necessary after oscillator replacement and is generally not

 

recommended.

 

 

Related Performance Test

10 MHz Reference Accuracy

Description

The frequency of the internal 10 MHz frequency reference is compared to a known frequency standard and adjusted for minimum frequency error. This procedure does not adjust the short-term stability or long-term stability of the A21 10 MHz ovenized crystal oscillator (OCXO). Stability is determined by the characteristics of the particular oscillator and the environmental and warmup conditions to which it has been recently exposed. The spectrum analyzer must be on continuously for at least 24 hours immediately prior to oscillator adjustment to allow both the temperature and frequency of the oscillator to stabilize.

2-48 Adjustment Procedures

Page 88
Image 88
Agilent Technologies 856290216 service manual MHz Reference Adjustment-OCXO Non-Option