RAM Check Errors (711 to 716)

The instrument power-on diagnostics check the program RAM. This includes the two RAMS used for STATE storage. If any STATE information is found to be invalid, all data in that RAM is destroyed. A separate error code is generated for each defective program RAM. All RAM is battery-backed. See “State- and Trace-Storage Problems” in Chapter 10, “Controller Section.”

711 RAM U303 Checksum error of system RAM A2U303.

712 RAM U302 Checksum error of system RAM A2U302.

713 RAM U301 Checksum error of system RAM A2U301.

714 RAM U300 Checksum error of system RAM A2U300.

715 RAM U305 Checksum error of system RAM A2U305.

716 RAM U304 Checksum error of system RAM A2U304.

Microprocessor Error (7 17)

717 BAD UP!!

Microprocessor not fully operational. Refer to Chapter 10, “Controller

 

Section.”

Battery Problem (718)

If STATE or TRACE data is found to be corrupt, the processor tests the display RAMS and the program RAMS containing the STATE information. If the RAMS are working properly, this error message is generated. To check the BTl Battery and the battery backup circuitry, refer to “STATE and TRACE Storage Problems” in Chapter 10, “Controller Section.”

718 BATTERY? Nonvolatile RAM not working; check battery BTl. This error can also be generated if the battery has been disconnected then reconnected. If this is the cause, cycling power clears the error.

Model Number Error (719)

If this error occurs, return the instrument to a service center for repair.

719 MODEL #? Could not read ID string from EEROM A2U501.

General Troubleshooting 7-39