Agilent Technologies 856290216 System Errors 600 to, Digital and Checksum Errors 700 to

Models: 856290216

1 482
Download 482 pages 3.94 Kb
Page 256
Image 256

583

RBW

30K

Unable to adjust 30 kHz resolution bandwidth.

584 RBW IOOK

Unable to adjust 100 kHz resolution bandwidth.

585

RBW

300K

Unable to adjust 300 kHz resolution bandwidth.

586 RBW IM

Unable to adjust 1 MHz resolution bandwidth.

587

RBW

30K

Unable to adjust 30 kHz resolution bandwidth.

588 RBW IOOK

Unable to adjust 100 kHz resolution bandwidth.

589

RBW

300K

Unable to adjust 300 kHz resolution bandwidth.

590 RBW IM

Unable to adjust 1 MHz resolution bandwidth.

591

LOG

AMPL

Unable to adjust amplitude of log scale.

592

LOG

AMPL

Unable to adjust amplitude of log scale.

593

LOG

TUNE

Limiter calibration tune error from DC logger calibration.

594

LOG

OFST

Attenuator calibration offset error from DC logger calibration.

595 LOG ATTN

Attenuator calibration absolute error from DC logger calibration.

596

LOG

FID

Fidelity error from DC logger calibration.

597

LOG

OFST

Fidelity offset error from DC logger calibration.

598

LOG

OFST

Fidelity offset unstable from DC logger calibration.

599

LOG

GAIN

Fidelity gain error from DC logger calibration.

System Errors (600 to 651)

ADC timeout errors occur if the A2 controller assembly frequency counter is faulty. Refer to Chapter 8, “ADC/Interface Section.”

600 SYSTEM Hardware/firmware interaction; check other errors.

601 SYSTEM Hardware/firmware interaction; check other errors.

650 OUTOF RG ADC input is outside of the ADC range.

651 NO IRQ Microprocessor is not receiving interrupt from ADC.

Digital and Checksum Errors (700 to 799)

EEROM Checksum Errors (700 to 704)

Faults on the A2 controller assembly can cause these errors. Refer to Chapter 10, “Controller Section.” Although some of these errors might result in a blanked display, it is possible to read these errors over HP-IB. Refer to “Troubleshooting to a Functional Section” in this chapter.

The EEROM on A2 is used to store data for frequency response correction, elapsed time, focus, and intensity levels. Error codes from 700 to 703 indicate that some part of the data in EEROM is invalid. An EEROM error could result from either a defective EEROM or an

General Troubleshooting 7-37

Page 256
Image 256