Procedure 13. Rear Frame/Rear Dress Panel

4.Place the coax cable BNC connectors into the appropriate rear frame holes as described below. Use a 9/16-inch nut driver to attach the dress nuts holding the BNC connectors to the rear frame.

Rear Panel Jack

RF Cable 1

W24, coax 5 W23, coax 93

W25, coax 4

W18, coax 97 W31, coax 8 W58, coax 0

5.Secure the rear frame to the spectrum analyzer main deck, using four panhead screws (1). See Figure 4-22.

6.Secure the rear frame to the spectrum analyzer side frames using three flathead screws per side. Use caution to avoid damaging any coaxial cables.

7.Place the spectrum analyzer top-side-up on the work bench.

8.Pull the red and black battery wires through the rear frame battery-assembly hole. Solder the red wire to the positive lug of the battery-assembly, and the black wire to the negative lug. Replace the battery.

9.Secure the battery assembly to the rear frame, using two flathead screws.

10.Connect the fan and line-power cables to A6J3 and A6JlOl on the A6 power supply.

11.Snap the A6AlW3 post-accelerator cable to the CRT assembly.

12.Snap the black grommet protecting A6AlW3 into the CRT shield.

13.Ensure that all cables are safely routed and will not be damaged when securing the A6 cover.

14.Secure the power-supply cover shield to the power supply, using three flathead screws (1). One end of the cover fits into a slot provided in the rear frame assembly. Ensure that the extended portion of the cover shield is seated in the shield wall groove. See Figure 4-23.

15.Connect the HP-IB cable to A2J5.

16.Fold the A2, A3, A4, and A5 assemblies into the spectrum analyzer as described in procedure 5.

4-46 Assembly Replacement

Page 171
Image 171
Agilent Technologies 856290216 service manual W24, coax 5 W23, coax