Single Test

Press Single Test to run the adjustment indicated by the pointer. Once the adjustment is

running, press Restart to abort and restart the test.

Calibrate Power Sensor

The frequency response adjustment program keeps track of which power sensor is being used and the elapsed time since it was last calibrated.

The program prompts the user to recalibrate the power sensor if more than 2 hours elapses since the last calibration. Also, if the power sensor is changed, the new power sensor must be calibrated.

If there is a significant change in ambient temperature, or improved power meter accuracy is desired, it is advisable to recalibrate the sensor more often than the program requires. Press Cal Sensor and follow the instructions on the computer screen to recalibrate the power sensor.

List Equipment

To obtain a list of required test equipment for running a test, move the pointer to the test, press List Equip. All HP-IB controlled equipment and passive devices, other than required cables and adapters, are listed. If a test is flagged MISSING ETE but all test equipment appears present, press List Equip to see what is needed, then return to the Conditions Menu and

verify that the equipment is present. Press Cond Menu to return to the Conditions Menu from the Test Menu.

Frequency Response Adjustment Software 3-11