improper sequence of storing data in EEROM. Check the EEROM with the following steps:

1.Place the WR PROT/WR ENA jumper on the A2 controller assembly in the WR ENA position.

2.On the spectrum analyzer, press a MORE i SF 2,

SERJfICE CAL DATA, FLATNESS ,and FLATNESS DATA. Enter a value of 130. Press PREV MENU, STORE DATA, YES, and (DISPLAY).

3.Press INTEN, enter an intensity value of 90, and press STfHE INTEZ .

4.Press MQRE i OF 2 FOCUS, enter a focus value of 128, and press

STORE FCKUS. Turn the (LINE) switch off, then on, cycling the power.

5.If errors are still present, the EEROM A2U501 is defective. Refer to the EEROM replacement procedure in Chapter 5, “Replaceable Parts.”

700 EEROM Checksum error of EEROM A2U501.

701 AMPL CAL Checksum error of frequency response correction data.

702 ELAP TIM Checksum error of elapsed time data.

703 AMPL CAL Checksum error of frequency response correction data. Default values being used.

704 PRESELCT Checksum error of customer preselector peak data. External preselector data recalled in internal mode, or internal preselector data recalled in external mode. To clear the error, press (X), MBRE 1 OF 2, FACTORY PRSEL PK ,(SAVE),and SAVE PRSEL PK.

Program ROM Checksum Errors (705 to 710)

705 ROM U306

706 ROM U307

707 ROM U308

708 ROM U309

709 ROM U310

710 ROM U311

The instrument power-on diagnostics perform a checksum on each programmed ROM (A2 controller assembly). If an invalid checksum is found for a particular ROM, an error code is generated. If a defective programmed ROM is found, replace it with another ROM with the same HP part number. Refer to Chapter 5, “Replaceable Parts.”

Although some of these errors might result in a blanked display, it is possible to read these errors over HP-IB. Refer to “Troubleshooting to a Functional Section” in this chapter.

Checksum error of program ROM A2U306.

Checksum error of program ROM A2U307.

Checksum error of program ROM A2U308.

Checksum error of program ROM A2U309.

Checksum error of program ROM A2U310.

Checksum error of program ROM A2U311.

7-36 General Troubleshooting

Page 257
Image 257
Agilent Technologies 856290216 ROM U306 ROM U307 ROM U308 ROM U309 ROM U310 ROM U311, Program ROM Checksum Errors 705 to