1.Disconnect the line-power cord, remove any adapters from the front panel connectors, and place the spectrum analyzer on its front panel.

2.If an HP 85620A Mass Memory Module or HP 85629B Test and Adjustment Module is mounted on the rear panel, remove it. Loosen (but do not remove) the four rear-bumper screws, using a 4 mm hex wrench. Pull the cover assembly off towards the rear of the instrument.

Caution

When replacing the spectrum analyzer cover, use caution to avoid damaging

 

any cables.

 

 

3.When installing the cover assembly, be sure to locate the cover air vent holes on the bottom side of the spectrum analyzer. Attach with the four screws loosened in step 2, and tighten the four screws gradually to ensure that the cover is seated in the front frame gasket groove.

4.Torque each screw to 40 to 50 inch-pounds to ensure proper gasket compression to minimize EMI.

A

A l 5

SK122

 

 

 

Figure 4-1. Hinged Assemblies

4-4 Assembly Replacement