Table l-4. Recommended Test Equipment

Instrument

sources

jynthesized sweeper

(two required)

;ynthesizer/level generator

iynthesized signal generator

‘&e/function generator

LM/FM signal generator

Critical Specifications for Equipment Substitution

Frequency range:

10 MHz to 13.2 GHz

Frequency accuracy (CW): x l0-g/day

Leveling modes: Internal & External

Modulation modes: AM & Pulse

Power level range: -80 to +16 dBm

Frequency range: 200 Hz to 80 MHz

Frequency accuracy: 1 x 10e7/month

Flatness: f0.15 dB

Attenuator accuracy: <f0.09 dB External 10 MHz reference input Frequency resolution: 1 Hz

Frequency range: 100 kHz to 2.5 GHz Residual SSB phase noise at 1 GHz:

<-73 dBc/Hz at 10 Hz offset <-lo7 dBc/Hz at 1 kHz offset <-124 dBc/Hz at 10 kHz offset <-124 dBc/Hz at 100 kHz offset

Frequency range: 10 kHz to 50 MHz

Pulse width: 200 ns

3utput amplitude: 5 V peak-to-peak

Functions: pulse & triangle

Pulse rise time: <lOO ns

I’TL sync output

Frequency range: 1 MHz to 200 MHz Frequency modulation mode

Modulation oscillator frequency: 1 kHz FM peak deviation: 5 kHz

Xecommendec

Model

HP 8340A/B* HP 83630A opt 001, 008

HP 3335A+

HP 8663A

HP 8116A

HP 8640B

HP 8642A

Use

PAT,

WV

PAT,

WV

p,v

P,A

A

Part of microwave workstation

= performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting; J = operation verification

l-10 General Information

Page 33
Image 33
Agilent Technologies 856290216 service manual Table l-4. Recommended Test Equipment, Instrument Sources