Section 4 System Maintenance

CAUTION

Hot surfaces may cause burns. The UV lamp and nearby sur- faces are hot.

To remove the lamp:

1.Remove the two screws that secure the waste port mount- ing plate (Figure 4-5). The optical assembly is located behind this plate.

Figure 4-5 Waste port mounting plate

Lamp Mounting Screws

Twist Lock Clamp

Figure 4-6 Deuterium lamp location

2.Disconnect the wiring connector.

3.Release the twist lock cable clamp.

4.Remove the three lamp mounting screws.

5.Carefully pull the lamp straight out from the optical assembly.