Section 4 System Maintenance

WARNING

Risk of damage to eyesight and skin. Wear ultraviolet (UV) light eye protection and protect skin from exposure if the deuterium lamp is operating.

CAUTION

Hot surfaces may cause burns. The UV lamp and nearby sur- faces are hot.

To reset the circuit breaker, separate the pump and column modules (see section 2). Locate the circuit breaker illustrated in Figure 4-7 and press the button.

Circuit Breaker

Figure 4-7 Circuit breaker location

The thermal breaker is located behind the waste port panel (Figure 4-6). To reset the breaker, open the panel and locate the reset switch (Figure 4-8). Insert a small flat blade screwdriver into the opening and push the switch in to the right.

Thermal Breaker

Figure 4-8 Thermal breaker location