BioOptix™ 10
Section 2 Installation
2-17
Figure 2-20 BioOptix 10 Manual Hardware Control
window
9. In the Columns sectio n of this window, check each channel
number that should be primed. Configure the remaining
settings to match those shown in Figure 2-20.
10. Cl ick the Start Pump button. Observe the following for each
enabled channel:
Buffer B solution movement should be visible at the
inlet ports on the left side of the BioOptix 10.
After passing through the internal pump and valves,
Buffer B solution should be visible at the Pump Purge
port.
11. Continue to manually pump Buffer B solution through the
system and inspect the tubing connections.
If any connections are leaking, use a wrench to tighten the
fitting 1/4 turn, then check again for leaks. Repeat as neces-
sary.
If no leaks are found, proceed with step 12.
12. Cl ick the Stop Pump button.
13. Change the %B setting to 0 percent.
14. Cl ick the Start Pump button. Monitor the flow of Buffer A
through the system. Correct leaks as needed.
15. Cl ick the Stop Pump button.
16. Cl ick the Done button.
The system is ready for operation (Section 3).