MODEL 6081-P pH/ORP

SECTION 2.0

 

INSTALLATION

SECTION 2.0

INSTALLATION

2.1Considerations

2.2Unpacking and Inspection

2.3Pre-Installation Set Up

2.4Mechanical Installation

2.5Ground the Transmitter

2.6Power Module Installation

2.1 CONSIDERATIONS

The transmitter can be commissioned before or after installation. It may be useful to commission it on the bench, before installation, to ensure proper operation and to become familiar with its functionality. When applicable, make sure the instru- ments are installed in accordance with intrinsically safe or non-incendive field wiring practices. The device will be powered whenever the power module is installed. To avoid depleting the power module, make sure it is removed when the device is not in use.

Power module

The Model 6081 is battery powered. The power module with the wireless unit contains 2 “C” size primary lithium/thionyl chloride batteries. Each power module contains approximately .5 grams of lithium. Under normal conditions, the power module materials are self-contained and are not reactive as long as the power module integrity is maintained. Care should be taken to prevent thermal, electrical or mechanical damage. Contacts should be protected to prevent premature discharge. Use caution when handling the power module. The power module may be damaged if dropped from heights in excess of 20 feet.

Sensor

Make sensor connections through the cable entry in the enclosure. Be sure to provide adequate clearance for cover removal.

Environmental

Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.

2.2 UNPACKING AND INSPECTION

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, remove the transmitter. Be sure all items shown on the packing list are present. If items are missing, immediately notify Rosemount Analytical.

Save the shipping container and packaging. They can be reused if it is later necessary to return the transmitter to the factory.

5

Page 15
Image 15
Emerson 6081-P instruction manual Section Installation, Considerations, Unpacking and Inspection