MODEL | SECTION 10.0 |
| CALIBRATION – PH |
SECTION 10.0
CALIBRATION — pH AND ORP
10.1Introduction
10.2Procedure – Auto Buffer Calibration
10.3Procedure – Manual
10.4Procedure – Standardization
10.5Procedure – Entering a Known Slope Value
10.6ORP Calibration
10.1INTRODUCTION
For pH sensors,
The ORP calibration is a
A new pH sensor must be calibrated before use. Regular recalibration is also necessary.
A pH measurement cell (pH sensor and the solution to be measured) can be pictured as a battery with an extreme- ly high internal resistance. The voltage of the battery depends on the pH of the solution. The pH meter, which is basically a voltmeter with a very high input impedance, measures the cell voltage and calculates pH using a con- version factor. The actual value of the
In automatic calibration the transmitter recognizes the buffer and uses
| pH at 25°C | Standard(s) | |
| (nominal pH) |
|
|
| 1.68 | NIST, DIN 19266, JSI 8802, BSI (see note 1) |
|
| 3.56 | NIST, BSI |
|
| 3.78 | NIST |
|
| 4.01 | NIST, DIN 19266, JSI 8802, BSI |
|
| 6.86 | NIST, DIN 19266, JSI 8802, BSI |
|
| 7.00 | (see note 2) |
|
| 7.41 | NIST |
|
| 9.18 | NIST, DIN 19266, JSI 8802, BSI |
|
| 10.01 | NIST, JSI 8802, BSI |
|
12.45 | NIST, DIN 19266 | ||
|
|
|
|
Note 1: NIST is National Institute of Standards, DIN is Deutsche Institute für Normung, JSI is Japan Standards Institute, and BSI is British Standards Institute.
Note 2: pH 7 buffer is not a standard buffer. It is a popular commercial buffer in the United States.
35