MODEL 6081-P pH/ORP

SECTION 2.0

 

INSTALLATION

4.With gentle pressure, insert the keyed connector on the power module into the receptacle (labeled Power Module Connection on the drawing). The power module seats in the connection receptacle with an o-ring.

5.Confirm that the power module is fully inserted in the receptacle and properly aligned with the surrounding terminal block.

6.Replace the rear cover of the Model 6081 with the two screws to secure it to the central housing. Tighten screws and verify operation. Correct installation the rear cover will ensure that the power modules properly secured to power the transmitter.

Figure 2-6 .Securing the rear cover

To remove the power module, reverse the installation steps.

Note: A damaged or degraded o-ring may compromise the NEMA 4X/IP66 rating of the unit even when the rear cover is correctly installed. Please take care to protect the o-ring when removing and replacing the rear cover of the transmitter.