MODEL 6081-P pH/ORP

SECTION 8.0

 

PROGRAMMING THE TRANSMITTER

8.5 SETTING A SECURITY CODE

8.5.1 Purpose

This section describes how to set a security code. There are three levels of security: a. A user can view the default display and information screens only.

b. A user has access to the calibration menus only. c. A user has access to all menus.

The security code is a three-digit number. The table shows what happens when security codes are assigned to Calib (calibration) and Config (configure). In the table XXX and YYY are the assigned security codes.

 

Code assignments

 

 

 

Calib

Config

What happens

 

 

000

XXX

User enters XXX and has access to all menus.

 

 

XXX

YYY

User enters XXX and has access to calibration menus only. User enters YYY and has access to

 

 

 

all menus.

 

 

XXX

000

User needs no security code to have access to all menus.

 

000

000

User needs no security code to have access to all menus.

 

 

 

 

 

8.5.2 Procedure: Setting a security code.

1.Press MENU. The menu screen appears. Choose Program.

2.Choose >>.

3.Choose Security.

4.Choose Calib or Config.

a.If you chose Calib, enter a three-digit security code.

b.If you chose Config, enter a three-digit security code.

5.To return to the main display, press MENU the EXIT.

29

Page 39
Image 39
Emerson 6081-P instruction manual Setting a Security Code, Procedure Setting a security code, Choose Calib or Config