MODEL | SECTION 10.0 |
| CALIBRATION – PH |
During automatic calibration, the transmitter also measures noise and drift and does not accept calibration data until read- ings are stable. Calibration data will be accepted as soon as the pH reading is constant to within the
In manual calibration, the user judges when pH readings are stable. He also has to look up the pH of the buffer at the temperature it is being used and enter the value in the transmitter.
Once the transmitter completes the calibration, it calculates the calibration slope and offset. The slope is reported as the slope at 25°C. Figure
The transmitter can also be standardized. Standardization is the process of forcing the transmitter reading to match the reading from a second pH instrument. Standardization is sometimes called a
FIGURE 10-1. Calibration Slope and Offset
36