MODEL | SECTION 10.0 |
| CALIBRATION – PH |
10.6 ORP CALIBRATION
10.6.1 Purpose
1.For process control, it is often important to make the measured ORP agree with the ORP of a standard solu- tion.
2.During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.
10.6.2 Preparation of ORP standard solutions
ASTM
Iron (II) - iron (III) standard is available from Rosemount Analytical as PN
10.6.3 Procedure
1.Press MENU. The main menu screen appears. Choose Calibrate.
2.Choose ORP.
3.The top line shows the actual ORP or redox potential (Live). Once the reading is stable, change the number in the second line to the desired value. Press ENTER.
4.The screen on the left will appear briefly.
5.The display returns to the Cal Sensor screen. Press EXIT. Choose the other sensor and repeat steps 2 through 4.
41