NGA 2000 CLD

Instruction Manual

HAS60E-IM-HW

June 2007

WARNING

Before opening gas paths they must be purged with ambient air or neutral gas (N2) to avoid hazards caused by toxic, flammable, explosive or harmful to health sample gas components!

WARNING

INTERNAL UV SOURCE !

Ultraviolet light from the ozone generator can cause permanent eye damage ! Do not look directly at the ultraviolet source !

Use of ultraviolet filtering glases is recommended.

WARNING

TOXIC CHEMICAL HAZARD!

The optional UV lamp contains mercury. Lamp breakage could result in mercury exposure ! Mercury is highly toxic if absorbed through the skin or ingested, or if vapors are inhaled.

Handle lamp assembly with extreme care. If the lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.

Immediately clean up and dispose of the mercury spill and lamp residue as follows:

Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump and aspirator bottle with a long capillary tube. (Alternately, a commercially available mercury spill clean-up kit is recommended.)

Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all mercury, lamp residue and debris into a platic bottle which cab be tightly capped.

Label and return to hazardous material reclamation center. Do not place in the trash, incinerate or flush down the sewer.

Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sulfur dust.

P - 6 Preface

Emerson Process Management GmbH & Co.OHG