NGA 2000 CLD
Instruction Manual
June 2007
Leak Test
The CLD Analyzer Module is completely tested at the factory for gas leakage (leakage rate - 7.5 hPa/min. with He or - 2.5 hPa/min. with N2). The user is responsible for leakage testing at the inlet and outlet fittings on the rear panel minimum twice a year (see Section
NOTE
Electrical installation must be in compliance with the requirements of NAMUR and DIN VDE and/or any applicable national or local codes (like local electricity supply enterprises (ESE)).
Contaminants
White crystal deposits on the windows of the reaction chamber, and the clogging of capillaries and vent are usually due to sample contaminates such as ammonia reacting with the high ozone levels and NO components.
To eliminate the contaminates, the sampling system should be reworked or a preventive maintenance program developed (if dropout is not excessive). Another source of crystalline formation is contaminated air.
NOTE
Contamination is a result of a not properly working sample handling systrem and is not be covered by Emerson Process Management warranty.
Refer to figure
1.Interconnection with the Backplane of the Plat- form. (See Instruction Manual for the NGA 2000 Platform).
2.
Connect Analyzer Module POWER to a 24 VDC, 10 A power source, either the Platform or external power source.
Regulator Adjust Slot | Metering Valve |
|
| Network |
FLOW | LON | Connection “1” |
| 1 |
|
| LON |
SAMPLE | 2 |
BALANCE |
Polarity
1 +
24 V 2 -
3 GND
Network Connection “2”
Power
Connection
REGULATOR
T 6A
~
250 V
Fuse
Figure
2 - 4 Installation | Emerson Process Management GmbH & Co.OHG |