Instruction Manual
June 2007
NGA 2000 CLD
b.Serial interface Setup
The submenu "Serial interface Setup" is used to set the parameters for data transfer between the analyzer and external devices. The choices in this menu depend on the configuration of the analyzer. The full specification of the serial interface is described in its own manual.
In the System SIO Module menu (Figure
Line#1 | 0.000 ppm |
| |
|
|
Baud rate: | 19200 |
Data bits: | 8 |
Stop bits: | 1 |
Parity: | None |
Echo mode: | Disabled |
Handshake: | Xon/Xoff |
Transmission delay: | 0 |
Type of installed serial interface: | RS232 |
Communication protocol: | AK |
Special protocol definitions... |
|
Pressure limits... |
|
Measure | Back... |
Line#1 | 0.000 ppm | |
| ||
|
| |
Device address | 1 | |
|
|
| Measure |
| Back... |
| More... |
|
|
|
Figure
The value can range from 1 to 50.
Figure
Options:
•Baud rate:
300, 1200, 2400, 4800, 9600, 19200
•Data bits: 7, 8
•Stop bits: 1, 2
•Parity: None, Even, Odd
•Echo mode: Enabled, Disabled
•Handshake: None, Xon, Xoff
•Transmission delay: 0...100.
Type of installed serial interface: RS232, RS485/2w, RS485/4w, RS485/4w bus, None.
Communication protocol: AK, MODBUS RTU, None (not applicable to CLD)
NOTE
The "special protocol definitions..." line accesses a submenu for setting the parameters of the AK and MODBUS RTU (not available yet) communication protocols.
Emerson Process Management GmbH & Co.OHG | Operation 3 - 11 |