NGA 2000 CLD
Instruction Manual
June 2007
It is the responsibility of the user to measure efficiency of the
The reactant material used in the converter provides the optimum combination of high conversion efficiency and low ammonia interference. Unlike most competitive ana- lyzers, the NGA 2000 CLD Analyzer Module utilizes a reactant material that gradually becomes more efficient at a given temperature. Thus, after a period of use, operation at a lower temperature setpoint than initially required is possible.
a.Test Setup for Measurement of Conversion Efficiency
A typical setup for measurement of conversion efficiency is shown in figure
A cylinder on nitric oxide standard gas consisting of NO in N2.
The concentration on NO in the standard gas should be about the
An ozone generator utilizing an ultraviolet lamp, not a corona discharge.
A corona discharge ozone generator is undesirable because it may produce oxygen atoms, which can then combine with atmospheric nitrogen to form NO. The result can erroneuosly high value for the measured conversion efficiency.
b.Test Procedure
1.Measure converter temperature in the
Note present reading as a reference for compari- son with subsequent readings.
2.Lower converter temperature to 300 °C by setting control parameters, and wait 15 minutes for tem- perature equilibration.
3.Connect a Converter Efficiency Tester (e.g. Model 958) to the CLD Module (see figure
4.Attach the NO/N2 supply to C2, the air supply to C1, and the CLD Module inlet fitting to C3.
5.With the variable transformer off, switch the CLD Module to NO mode, and close valve MV1.
6.Open valve MV2 until the CLD Module SAMPLE Pressure Gauge reaches operating pressure and the BYPASS flowmeter indicates some bypass flow. Wait until stable readings are obtained by the CLD Module.
7.Zero and span the Analyzer output to indicate the value of the NO concentration being used. This value should be about 80 % of
Record this concentration.
8.Open valve MV1 (air supply metering valve) and adjust to blend enough air to lower the NO con- centration (as noted above) about 10 %.
Record this concentration.
9.Power up the ozonator, and increase its supply voltage until NO concentration noted in Step 8 is reduced to about 20 percent of the concentration
noted in Step 7. Wait for stabilization. NO2 is now being formed from the NO+O3 reaction. There must always be at least 10 percent
10.Switch the CLD Module to NOx mode. Total NOx concentration is now output to the network for dis- play. Record this concentration.
11.Turn off the ozonator, and allow the Analyzer read-
ing to stabilize. Total NOx concentration of the di- lute NO span gas initially used is displayed. Record this concentration.
3 - 14 Operation | Emerson Process Management GmbH & Co.OHG |