2-3 LOCATION

Install the CLD Analyzer Module in a clean, weather-pro- tected, vibration-free location free from extreme temperature variations and moisture. For best results, install the instrument near the sample stream to minimize sample transport time.

The analyzer should be installed within ±15° of horizon- tal.

NOTE

Unrestricted air flow to the rear of the Analyzer Module is critical to its performance and reliablilty.

2-4 GASES

a.Gas Conditioning (Sample Handling)

NOTE

All gases must be conditioned before supplying!

The gases must be

-dry

-free of dust (filtered for particulates down to two microns)

-free of aggressive components affecting gas paths materials (e.g. by corrosion)

-free of Ammonia etc. in order to prevent crystallin formation

Operating ambient temperature is 0 °C to 45 °C (32 °F to 113 °F). Temperature change should not exceed 10 °C (18 °F) per hour. The same temperature range restrictions apply to the location of the air and span gas cylin- ders.

before entering the Analyzer Module.

The gases should have a dew point 5 °C (13 °F) below coldest ambient temperature.

CAUTION

EXTERNALLY RESTRICT SAMPLE/CALIBRATION

GAS FLOW TO LESS THAN 2,200 cc/min.

Damage to internal components may occur if this flow level is exceeded.

No restrictor is provided in the sample inlet of this module. For those users who cannot externally limit sample flow, contact your local service or sales office.

MAXIMUM NOx

GAS SUPPLIED TO

SETTING ON OZONE

SETTING ON SAMPLE

LEVEL IN PARTS

REAR PANEL AIR

PRESSURE GAUGE

PRESSURE GAUGE

PER MILLION

INLET

(pressure values: gauge)

(pressure values: gauge)

 

 

 

 

 

 

690 hPa (10 psig); provides flow

344 hPa (5 psig); provides flow

800

Air

of approximately 500 cc/min. to

of approximately 200 cc/min. to

 

 

ozone generator

reaction chamber

 

 

 

 

 

 

690 hPa (10 psig); provides flow

344 hPa (5 psig); provides flow

2,500

Air

of approximately 1,000 cc/min. to

of approximately 70 cc/min. to

 

 

ozone generator.

reaction chamber.

 

 

 

 

 

 

1035 hPa (15 psig); provides flow

103 hPa (1.5 psig); procides

10,000

Oxygen

of approximately 1,000 cc/min to

flow of approximately 20 cc/min.

 

 

ozone generator.

to reaction chamber.

 

 

 

 

Table 2-1: Gas Specifications

2 - 2 Installation