NGA 2000 CLD

Instruction Manual

HAS60E-IM-HW

June 2007

4-6 CONVERTER

CAUTION

HIGH TEMPERATURES !

While working at thermostated components inside the Analyzer Modules hot components may be accessible!

Refer to figure 4-1and figure 4-2. To replace the converter or temperature sensor, disconnect the two pneumatic tubes and two electrical connections. Unlace the heater blanket, and remove the converter. Reassemble in reverse order, ensuring that the converter is oriented with the glass cloth at the bottom and the sensor is oriented correctly inside the heater jacket.

Converter complete (655250)

Heater Jacket (655228-R1)

Temperature Sensor

(655282-R1)

Wrap with

Converter Tube

Connectors

Aluminium Foil

(655227)

(632784)

Figure 4-2: Converter Assembly

4 - 4 Maintenance and Service

Emerson Process Management GmbH & Co.OHG