NGA 2000 CLD

If there is a leakage, then proceed as follows:

a)Using Helium (He) for test medium

Supply Analyzer Module with the test medium with a pressure of approx. 2,000 hPa (15 psig).

Look for the leak with a helium leak detector.

b)Using Nitrogen (N2) for test medium

Supply Analyzer Module with the test medium with a pressure of approx. 2,000 hPa (15 psig).

Liberally cover all fittings, seals, and other pos- sible sources of leakage with a suitable leak test liquid such as SNOOP. Bubbling or foam- ing indicates leakage.

Remove the source of leak.

Perform a leakage test once more as decribed obove.

.