NGA 2000 CLD
Instruction Manual
June 2007
After performing
In the following, analyzer function control and calibration procedure is described
CLD |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
7.50 |
|
|
|
|
|
|
|
|
|
|
|
| ppm NO | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
0.00 |
|
|
|
|
|
|
|
| Range 1 | 10.00 |
| |||||||||||
Sample Flow: | 1300 ml/min |
|
|
|
|
| ||||||||||||||||
200 |
|
| 1500 | |||||||||||||||||||
Sample press: |
|
| 4.0 hPa |
| 50.0 |
|
| 490.0 | ||||||||||||||
Ozonator: |
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
Converter temp: |
|
| 30.9 C |
| 150.0 | 500.0 |
Display Status...
Main...
Channel
BasicCal
Figure
After starting up the Analyzer Module, the Measure Mode Display is displayed as shown in figure
To now check back physical parameters of the CLD Analyzer Module with the values defined in your Test Data Sheet, that you received with the analyzer, you have to select the "Physical
Enter the diagnostics menu "Physical Measurements" as follows, using the softkeys F1 through F5:
Main (Menu)...
Expert Controls and Setup...
Analyzer Module Setup...
Physical Measurements...
Line#1 | Physical Measurements |
| 0.000 ppm | ||||||||
|
|
|
|
| 0.7 hPa | ||||||
Sample Capillary Pressure: |
|
|
|
|
| ||||||
Ozone Supply Pressure: |
|
|
|
|
| 4.0 hPa | |||||
Ozonator Status: |
|
|
| OFF - PRES. SW. | |||||||
Ozonator Power: |
|
|
|
|
| Enabled | |||||
Bypass Flow: |
|
|
|
|
|
| 2 ml/min | ||||
Converter temperature: |
|
|
|
|
| 320 C | |||||
Ozonator temperature: |
|
|
|
|
| 51.5 C | |||||
Sensor temperature: |
|
|
|
|
| 1.1 C | |||||
Block temperature: |
|
|
|
|
| 51.5 C | |||||
Capillary Flow Rate: |
|
|
|
| 199.9 ml/min | ||||||
| Pressure limits... |
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |
| HOME |
|
| ESCAPE |
| MORE |
| HISTORY |
| INFO |
The menu "Physical Measurements" lets you monitor physical measurement parameters of the CLD analyzer.
During
After warm up, check
•Block Temperature
•Detector Temperature
•Converter Temperature
with the operating values given by yout Test Data Sheet. You will find an excerpt of a Test Data Sheet as an example on the next page in figure
NOTE:
As long as operating temperatures are not yet reached by the internal components of the ana- lyzer, it is not ready for operation.
After
Supply gases at the pressures given in the Test Data Sheet and restrict flow by an external flow limiter to 1.3 through 1.5 l/min.
Internal sample gas and ozonator pressures:
You will find internal physical paramters in the Test Data Sheet that you have received with the ana- lyzer. Externally supply gases at the given pressures with an external adjustment.
Option Bypass Flow:
Depending on measuring capillary, bypass flow should by 900 - 2,000 cc/min. If no bypass is in- stalled, restrict flow externally to 1.3 - 1.5 l/min.
Figure