3-5 SYSTEM & NETWORK I/O MODULE

Output number:

CONTROLS (SETUP) - SYSTEM SIO

Choose the desired analog output (1-8) to set the pa-

To adjust SIO functions, select

rameters. The number of outputs depends on the ana-

Main (Menu)...

lyzer configuration as 2, 4, 6 or 8.

 

Analyzer and I/O, expert controls & setup...

System & network I/O module controls...

System SIO module...

Line#1

 

 

0.000 ppm

 

 

-- System SIO Module --

Analog output setup...

 

 

 

Serial interface setup...

 

 

 

Relay outputs setup...

 

 

 

Module installed...

 

 

YES

.

 

 

 

 

MEASURE

 

 

Back...

 

Figure 3-11: System SIO Module Menu

System SIO module menu allows you to adjust different SIO functions, like analog outputs and serial interfaces.

a.Analog Output Setup

In the System SIO Module menu (Figure 3-11) select "Analog output setup..."

 

Line#1

 

 

0.000 ppm

 

 

-- Analog Output Setup --

 

 

 

 

 

 

 

 

 

 

Output number:

 

 

1

 

 

Choose signal source module...

 

 

 

 

 

Choose signal...

 

 

0.00

 

Signal value for 0% output:

 

 

 

Signal value for 100% output:

 

 

100.00

 

Output current:

 

 

0...20 mA

 

Hold output during calibration:

 

 

No

 

Signal name:

 

 

????

 

Current signal value:

 

 

-10.00

 

Source module:

 

 

Line#1

 

 

 

 

 

 

 

Measure

 

Back...

 

More...

Figure 3-12: Analog Output Setup Menu

Choose signal source module...

Select the "Analyzer Modules" submenu by selecting the "Choose signal source module..." line and pressing the Return () softkey.

Select the tag of the desired reference channel with the

or softkeys and then press the or softkey. The display will return to the previous menu automatically and the selected reference channel will be displayed in the "Source module:" line.

The available selections may be different depending on the installed modules.

Line#1

0.000 ppm

-- Analyzer Modules --

1 Control Module: 0.0 1000723 Line#1: 1.0 CLD: 1.0 MLT/CH3

 

HOME

 

NO/NOx

 

ZERO

 

SPAN

 

INFO

 

 

 

 

 

Figure 3-13: Analyzer Modules Menu

Choose Signal...

Select the "Signals" submenu by selecting the "Choose Signal..." line and pressing the softkey. (The list of signals will depend on the module chosen.)

Press the F5 softkey to go to additional menus to choose the Primary Variable signal for the analog output. The Primary Variable is the actual NO or NOx concentration.

See Section 5.2.1.1 of CLD Software Manual for complete list of signals.

The signal chosen here will be applied to the analog output (1-8) chosen above.