NGA 2000 CLD

Instruction Manual

HAS60E-IM-HW

June 2007

c.Relay Outputs Setup

There are three relays on the SIO board. The contact logic can be set with a jumper on the SIO board to select NO (normally open) or NC (normally closed). Full details of the SIO board are contained in its own manual.

In the System SIO Module menu (Figure 3-11)select "Relay outputs setup..."

 

Line#1

 

 

0.000 ppm

 

 

 

-- Relay Outputs Setup --

 

 

 

 

 

 

 

 

 

Output number:

1

 

 

 

Invert signal:

 

 

Disabled

 

 

 

Choose source module...

 

 

 

 

 

 

Choose signal...

 

 

 

 

 

 

 

 

 

 

Signal comes from:

????

 

 

Signal name:

????

 

 

Actual status:

 

 

Off

 

 

 

 

 

 

 

 

Measure

 

 

Back...

 

 

 

Figure 3-20: Relay Outputs Setup Menu

Output number:

Corresponds to the relay number 1-3.

Invert signal:

"Disabled" signal is normal, "Enabled" signal is inverted.

Choose source module...

Line#1

0.000 ppm

-- Choose Source Module --

Control Module: 0.0

1000723 Line#1: 1.0

 

Measure

 

<<<

 

Back...

 

>>>

 

 

 

 

Choose signal...

Line#1

-- Choose Signal --

 

 

 

0.000 ppm

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

Function control

 

 

 

 

 

Maintenance request

 

 

 

 

 

 

 

 

Failure

 

 

 

 

 

 

Cal. In progress

 

 

 

 

 

 

Zero in progress

 

 

 

 

 

 

Span in progress

 

 

 

 

 

 

 

 

Zero failed

Pressure limits...

 

 

 

 

 

 

Span failed

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

<<<

 

...Back

 

 

>>>

Figure 3-22: Choose Signal Menu

Choose desired signal for the relay output number (1-3) being configured.

The list of signals will depend on the chosen module. If available, press the >>> (F5) softkey for additional sig- nals.

The three lines displayed at the bottom of the "Relay Outputs Setup" menu show the current status of the selected relay output.

Signal comes from: The module chosenfrom the "Choose Source Module" menu.

Signal name: The signal chosen from the "Choose Sig- nal" menu.

Actual status: The current status of the signal; Off or On.

Figure 3-21: Choose Source Module Menu

Choose desired source module for the relay output number (1-3) being configured.

The list of modules will depend on the installed modules.

3 - 12 Operation