NGA 2000 CLD

Instruction Manual

HAS60E-IM-HW

June 2007

b.Flow Balance Adjustment:

To adjust the Flow Balance Valve, select:

Main (Menu)...

Analyzer basic controls (calibration) & setup...

 

Line#1

 

 

Basic Controls

 

 

 

0.000 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement range number:

 

 

 

 

1

 

 

Range upper limit:

 

 

 

 

 

 

24.9 ppm

 

 

Automatic range change control:

 

 

 

Disabled

 

Ranges with valid calibration:

 

 

 

 

1&2

 

Calibration status:

 

 

 

 

 

 

READY

 

Span gas concentration:

 

 

 

 

 

 

10.0 ppm

 

Status:

 

 

 

 

 

 

 

 

NORMAL

 

Measurement Mode:

 

 

 

 

 

 

NOx mode

 

Ozonator status:

 

 

 

 

 

 

OFF -PRES. SW.

 

Ozonator power:

 

 

 

 

 

 

 

 

Enabled

 

Pressure limits...

 

 

 

 

 

 

 

 

 

 

 

HOME

 

 

NO/NOx

 

 

ZERO

 

SPAN

 

 

INFO

Span Adjustment

After that, do a Span adjust in the same measuring range.

For that, push the F4 softkey in the Basic Controls menu, so that the Analyzer Span menu turns up:

Line#1

0.000 ppm

 

Analyzer Span

Are you sure ?

 

You must have span gas flowing through the analyzer.

 

Calibration time:

 

 

 

 

24 s

 

Measurement range number:

 

1

 

Span ranges:

 

 

 

SEPARATELY

 

Calibration status:

 

 

READY

 

Error message for last span:

 

 

CAL OK

 

 

 

 

 

 

 

 

 

 

HOME

 

 

NO/NOx

 

SPAN

 

INFO

Figure 3-8: Basic Controls Menu

To adjust the flow balance, select the measuring range that is really in use and do a Zero- / Span adjust in NO mode.

Zero Adjustment

In the Basic Controls menu, push the F3 softkey and enter the Analyzer Zero menu.

Line#1

 

 

Analyzer Zero

 

0.000 ppm

 

 

 

 

 

 

 

Are you sure ?

 

 

 

 

 

 

You must have zero gas flowing through the analyzer.

 

 

Calibration time:

 

 

 

 

 

24 s

Measurement range number:

 

 

1

Zero ranges:

 

 

 

 

SEPARATELY

Calibration status:

 

 

 

READY

Error message for last zero:

 

 

 

CAL OK

.

 

 

 

 

 

 

 

 

HOME

 

 

ESCAPE

 

ZERO

 

 

INFO

Figure 3-9: Analyzer Zero Menu

Push the F3 softkey again, to start the zeroing process. Be sure to have zero gas flowing through your analyzer and to have it purged it from sample gas before.

When the zeroing process is finished, "Calibration sta- tus" will turn to "READY".

Figure 3-10: Analyzer Span Menu

Push the softkey F4 again to start the spanning sequence. Be sure to have span gas flowing through your analyzer and to have it purged before.

When the spanning sequence is finished, the "Calibra- tion status" will turn to "READY".

When having completed the zero- and span adjust in NO mode, switch over to NOx mode by pressing the F2 softkey in the Basic Controls menu. Observe the measuring value after switching into NOx mode:

If the measuring value stays constant (maybe af- ter a short peak immediately after switching into NOx mode), your flow balance is perfectly adjusted and your analyzer is ready for measurement.

If the measuring value changes after switching to NOx mode, turn the flow balance valve carefully until the difference between measuring values in NO mode and NOx mode is minimized.

To make sure that your analyzer is correctly calibrated, switch back to NO mode, if necessary, repeat zero- and span adjustment and check for an existing difference between measurement in NO mode and in NOx mode.

3 - 6 Operation

Emerson Process Management GmbH & Co.OHG