NGA 2000 CLD

Instruction Manual

HAS60E-IM-HW

June 2007

Output(s) value on analyzer failure:

Choose the desired signal level to cause a failure condi- tion. The choices are:

Acutal

Source module: The name of the module chosen in the "Choose signal source module" menu.

Pressing the F5 (More...) softkey changes to the submenu "Special Scaling for Concentration Signal"

BeginOfRange

EndOfRange

BeginOfRange-10%

BeginOfRange+10%

Output number:

Choose the output number (1-8) for setting the fine ad- justment.

Operation mode:

Line#1

0.000 ppm

 

-- Special Scaling for Concentration Signal --

 

(Scaling is the same as range limits)

 

 

Output #1:

Yes

Output #2:

Yes

Output #3:

Yes

Output #4:

Yes

Output #5:

Yes

Output #6:

Yes

Output #7:

Yes

Output #8:

Yes

 

Measure

 

Back...

 

More...

 

 

 

Normal: The absolute measurement signal will be sent to the analog output.

Adjust 0V: Used to set the display equal to the analog output for 0 V and 0 mA. Life zero signals (4 - 20 mA and 2 - 10 V) are set automatically and cannot be adjusted.

Adjust 10V: Used to set the display equal to the analog output for 10V und 20mA.

Select the "Fine adjustment for 0% output" and/or "Fine adjustment for 100 % output" lines with the or softkey. Adjust to the desired value with the or softkey and confirm with the softkey.

The range of values are:

3,000 to 6,000 for 0 % (default 4096)

600 to 1,000 for 100 % (default 819)

The last three lines of the "Analog Output Setup" menu are displayed only for configuration values of the analog output.

Signal name: The name of the signal chosen the "Choose signal" menu.

Current signal value: The current value of the variable.

Figure 3-16: Special Scaling for Concentration

Signal Menu

This menu allows for the setting of each of the 8 outputs to be the same as the range limits "Yes" or as to set on the previous menus.

Pressing the F5 (More...) softkey changes to the submenu "Analog Output Updates per Second."

Line#1

0.000 ppm

 

-- Analog Output Updates Per Second --

 

1

Output #1:

0

Output #2:

0

Output #3:

0

Output #4:

0

Output #5:

0

Output #6:

0

Output #7:

0

Output #8:

0

 

Measure

 

Back...

 

 

 

 

Figure 3-17: Analog Output Updates per

Second Menu.

This menu allows for the setting of the update rate for each of the 8 outputs.

3 - 10 Operation

Emerson Process Management GmbH & Co.OHG