
Instruction Manual
Model NGA2000 PMD
Install the Analyzer Module in a clean, non- hazardous, weather protected, vibration free location free from extreme temperature varia- tions. For best results, either install the module near the sample stream to minimize sample transport time or supply a flow greater than necessary and route only the appropriate amount through the Analyzer Module.
An
Sample Gas
Observing these requirements are critical. Note the following:
•Excessive vibration can cause a noisy readout. To minimize vibration effects, the detector/magnet assembly is envel- oped in a
•The user should ensure, when making any internal electrical connections, that no cables are placed in contact with the detector assembly or associated internal sample inlet and outlet tubing.
•Magnetic susceptibilities and partial pressures of gases vary with tempera- ture. Permissible ambient temperature range is 32°F to 113°F (0°C to 45°C).
•The interior of the Detector Assembly is maintained at approximately 144°F (62°C) by an electronically controlled heater. Prior to entering the detector assembly, the sample is heated in a coiled tubing to match the detector's temperature.
a.Requirements
Calibration Gases
Analyzer Module calibration requires the establishment of zero and span calibra- tion points. This requires a zero standard gas to set the zero point span gas to es- tablish a calibration point at or near the upper range limit.
Sample gas should be
Temperature
Sample temperature at the inlet should be from 50°F to 150°F (10°C to 66°C). A maximum entry temperature of 110°F
(43°C) is recommended to prevent cooling of the sample and possible internal condensation. Such condensation could damage some components of the Analyzer Module. This recommendation can be ignored if a thoroughly dry sample is examined.
Pressure
Sample exhaust pressure limits are
To prevent
The outlet port is commonly vented to the atmosphere. Any change in barometric pressure has a directly proportional effect on the indicated percent of oxygen, and should be neutralized through manual or computer correction of data. Note the following example:
Rosemount Analytical Inc. A Division of Emerson Process Management |