Instruction Manual

760007-A July 2003

Menu: 3 ZEROl1

Model NGA2000 PMD

Zero/Span Calibration help

This allow manual control of the zero and span. Flow zero gas, and make sure the gas value is correct: press the zero key to make the analyzer zero itself. Or select Edit measurement using zero offset, then scroll the reading with the up and down keys. In this way you can make the analyzer read what you want.

Then do the same with span gas. If the zero was not a real zero, the span action will change the zero reading; the last zero reading shows you what it would have been on the zero gas with the current span.

HOME

 

ESCAPE

 

MORE

 

INFO

 

 

 

Menu: 4 SPANl1

Span Calibration help

Use this screen to perform a span Calibration.

Either directly adjust the reading with the up and down arrow keys, or press the SPAN softkey to force a span calibration to the span gas concentration.

You should do a zero calibration before

HOME ESCAPE MORE BACK INFO

Menu: 5 FLOCHEKl1

Secondary Measurements help

This screen shows the auxiliary measurements made by the analyzer module the limits may be set by the user as warning alarms.

These readings are updated only when they change.

HOME

 

ESCAPE

 

INFO

 

 