Instruction Manual
Menu: 3 ZEROl1
Model NGA2000 PMD
Zero/Span Calibration help
This allow manual control of the zero and span. Flow zero gas, and make sure the gas value is correct: press the zero key to make the analyzer zero itself. Or select Edit measurement using zero offset, then scroll the reading with the up and down keys. In this way you can make the analyzer read what you want.
Then do the same with span gas. If the zero was not a real zero, the span action will change the zero reading; the last zero reading shows you what it would have been on the zero gas with the current span.
HOME |
| ESCAPE |
| MORE |
| INFO |
|
|
|
Menu: 4 SPANl1
Span Calibration help
Use this screen to perform a span Calibration.
Either directly adjust the reading with the up and down arrow keys, or press the SPAN softkey to force a span calibration to the span gas concentration.
You should do a zero calibration before
HOME 
ESCAPE 
MORE 
BACK 
INFO
Menu: 5 FLOCHEKl1
Secondary Measurements help
This screen shows the auxiliary measurements made by the analyzer module the limits may be set by the user as warning alarms.
These readings are updated only when they change.
HOME |
| ESCAPE |
| INFO |
|
|