Model NGA2000 PMD

Instruction Manual

760007-A July 2003

Range 1 (2, 3, 4) Factors. (Do steps 4 and 5 for each range.)

4.Select Zero Offset, press ENTER, adjust the value to 525000 with the up/down ar- row keys, and press ENTER. Do not press ESCAPE at any time unless reten- tion of prior settings is desired.

5.Select Span Factor, press ENTER, adjust the value to 0.000015 with the up/down arrow keys, and press ENTER. Do not

press ESCAPE unless retention of prior settings is desired.

6.Attempt to recalibrate the Analyzer Mod- ule according to the procedure outlined at the beginning of Section 3-5 on page 3-5. If re-calibration fails, return to the Range Factors menu, readjust factors, and at- tempt calibration again.

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-7