
Instruction Manual
Model NGA2000 PMD
Any gas having a composition other than 100% oxygen contains background gas, that is,
If the operator uses zero and span gases that contain the same background gas as the sample, calibration procedures automatically compensate. No adjustments are necessary.
If the background gas in the sample is different from that in the zero and/or span gases, the operator must take into consideration background effects to ensure correct readout. During entry of zero and span gas values in the Calibration Gas List, the instrument is not set to indicate the true oxygen content of the zero and span standard gases. It is set to indicate a slightly different value, relative to background gas, calculated to provide correct readout during subsequent analysis of sample gas.
Oxygen Equivalent Values of Gases
For computation of background corrections, the analyzer's response to each component of the sample must be known. Table
%O2 Equivalent of Gas = | Analyzer Response to Gas | X 100% | |
Analyzer Response to O2 | |||
|
|
For example, if the analyzer's response to oxygen is +100%, the response to xenon would be
The oxygen equivalent of a gas mixture is the sum of the contribution of the individual gas components.
Example: Zero Based Range
At lower range limit (i.e., 0% O2), composition of sample is: 80% CO2, 20% N2.
From Table
The percent oxygen equivalent of the mixture =
Rosemount Analytical Inc. A Division of Emerson Process Management |