Menu: 48 LINRANGE0l1
Set Linearity curve
The linearizer polynomials act over a certain range, not the same as the measurement range. The system uses the linearizer polynomial appropriate for
the measurement range chosen. This is the polynomial with the next higher linearizer range than the measurement range.
Polynomial coefficients may be edited for custom curves.
HOME |
| ESCAPE |
| INFO |
|
| |||
|
|
|
|
|
Menu: 49 PLMITSAl1
Pressure and flow Limits
These are settable limits on the sample
gas pressure and flow. The force alarms to occur and also act as end points on the bar graph display of their variable.
HOME |
| ESCAPE |
| INFO |
|
|
Menu: 50 CALFACTORS
Calibration Factors
Range 1 factors…
Range 2 factors…
Range 3 factors…
Range 4 factors…
Zero compensation factor: | 000.0 | |||
Span compensation factor: | 000.0 | |||
|
|
|
|
|