Menu: 114 CALFAIL

If it won’t calibrate…

Check that you are flowing the correct gas, and the gas concentration is what it is supposed to be. Make sure that the reading is stable before staring. If you have enabled or disabled the linearizer, you may have made it hard for the analyzer to calibrate. If so, go to the calibration parameters screen under Expert controls and set up, under Analyzer set up, and disable the limits checking. Recalibrate, and then enable the limits checking again.

If all else fails, manually adjust the calibration factors