Model NGA2000 PMD

Instruction Manual

760007-A July 2003

Menu: 102 RFACTORSIA

Range Factors

Shows the calibration factors for this range. Modify the zero factor for zero calibration, and the span factor for spanning this range. The take effect as soon as you press the enter key.

With zero gas, the zero factor should be the same as the raw reading.

RSTR MN restores the manufacturing values. RSTR ST restore the stored values.

HOME

 

ESCAPE

 

INFO

 

 

Menu: 103 RFHIST2A

Range 2 Factors

 

Manufacturer’s settings.

 

Zero offset:

000.0

Span factor:

000.0

Stored settings

 

Zero offset:

000.0

Span factor:

000.0

HOME

 

NEXT

 

RSTR MN

 

RSTR ST

 

INFO

 

 

 

 

Menu: 104 RFHIST3A

Range 3 Factors

 

Manufacturer’s settings.

 

Zero offset:

000.0

Span factor:

000.0

Stored settings

 

Zero offset:

000.0

Span factor:

000.0

HOME

 

NEXT

 

RSTR MN

 

RSTR ST

 

INFO

 

 

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendix A. Menu Displays

7-35