pressed range. Table 3-3 below shows recommended zero and span standard gases for

suppressed range oxygen indication in an Argon background.

RANGE % OXYGEN

RECOMMENDED ZERO

RECOMMENDED SPAN

STANDARD GAS

STANDARD GAS

 

0 to 1

Nitrogen

0.9% O2, balance N2

 

 

 

0 to 2.5

Nitrogen

2.3% O2, balance N2

 

 

 

0 to 5

Nitrogen

4.5% O2, balance N2

 

 

 

0 to 10

Nitrogen

9% O2, balance N2

 

 

 

0 to 25

Nitrogen

Air (20.93% O2)

 

 

 

0 to 50

Nitrogen

45% O2, balance N2

 

 

 

0 to 100

Nitrogen

100% O2

 

 

 

Table 3-2. Calibration Range for Various Zero Based Operating Ranges

RANGE % OXYGEN

RECOMMENDED ZERO STANDARD

RECOMMENDED SPAN

GAS

STANDARD GAS

 

50 to 100

50 - 50.5% O2 Balance AR

100% O2

 

 

 

70 – 100%

70 – 70.5% O2 Balance AR

100% O2

 

 

 

98 – 100%

98 – 98.5% O2 Balance AR

100% O2

 

 

 

99 – 100%

99 – 99.5% O2 Balance AR

100% O2

 

 

 

Table 3-3. Calibration Range for Various Suppressed Range Operations

If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), a possible solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module). Recalibrating with the appropriate gas(es) WILL NOT correct the problem because the ZERO OFFSET or SPAN FACTOR has been set to an extreme value in the proc- ess.

To remedy the problem, do the following:

1.Select the following from the Main Menu: Expert Controls and Setup (enter security

code if necessary), Analyzer Module Setup, and Calibration Parameters.

2.Using the down arrow, select Zero Ranges, press ENTER and, using the up/down arrows, toggle to SEPARATE. Do the same for the Span Ranges selec- tion. Do not press ESCAPE at any time unless retention of prior settings is de- sired.

3.Return to the Main Menu (HOME) and make the following selections: Expert Controls and Setup (enter security code if necessary), Expert Analyzer Controls, CAL softkey, FACTORS softkey, and