Instruction Manual
Model NGA2000 PMD
Menu: 105 RFHIST4A
Range 4 Factors |
|
Manufacturer’s settings. |
|
Zero offset: | 000.0 |
Span factor: | 000.0 |
Stored settings |
|
Zero offset: | 000.0 |
Span factor: | 000.0 |
HOME |
| NEXT |
| RSTR MN |
| RSTR ST |
| INFO |
|
|
|
|
Menu: 106 TWEAKI1
Midpoint correction set up
This function allows you to set up to three midpoints that the analyzer will
It does this with a
polynomial linearization. First disable the correction. Set the
Then enter the first midpoint gas value, run the gas, and when stable, press SET.
the actual reading, but the analyzer will
HOME |
| ESCAPE |
| MORE |
| INFO |
|
|
|
Menu: 107 ANALSETI3
Midpoint correction set up
Then go to the second set point, and repeat. You can use up to three midpoints.
When you are done, set the correction to
WARNING: make sure that you do not have excessive corrections. If the correction is too odd, the calibration routine will fail, and you will not be able to calibrate the analyzer. In this case, try it again.
You can perform this correction individually for each range.
HOME |
| ESCAPE |
| BACK |
| INFO |
|
|
|
Appendix A. Menu Displays | Rosemount Analytical Inc. A Division of Emerson Process Management |