Menu: 79 POLYSETI1

Polynomial set up

Select the range to linearizer.

Make sure that the span gas value is correct. Choose whether to define the gas concentrations as absolute values or as a percent to the span gas. Use percent if you are diluting the span gas for this. Get into the gas concentrations screens, and set the concentration for as many points as you want.

At each point, flow the gas of the correct value, and when the reading is stable, press

The analyzer will store the gas value and the reading

HOME

 

ESCAPE

 

MORE

 

INFO

 

 

 