Model NGA2000 PMD

Instruction Manual

760007-A July 2003

WARNING

OVER-VOLTAGE SPIKING

If this Analyzer Module is used with a non-Rosemount Analytical power supply, adding Rosemount Analytical PN 90331 Current Protector in series with the 24 V positive line will prevent over-voltage spiking and resultant fuse flowing when powering up the instrument.

CAUTION

HAND INJURY HAZARD

Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.

CAUTION

OVERBALANCE HAZARD

This Analyzer Module may tip instrument over if it is pulled out too far and the Platform is not properly supported.

NOTICE

Software compatibility is necessary for all NGA2000 components in your system to work together. The version of your Platform's software must be equal to or greater that the version of any other module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to order software upgrade kit PN 657150 for the Platform.

You can locate the version of each NGA2000 component as follows:

Platform Controller Board

Turn power ON.

The display will show "Control Module V2. ...". This is the software version.

Analyzer Module

Located on the right side of the Analyzer Module case.

I/O Module

Located on the backplane connector of the module. If no label is present, the module is Version 2.0.