Menu: 108 ACALSETI1A

Calibration Parameters

Disable the limits to recover from calibration failure Calibration averaging time sets the time used by the analyzer to average its reading. Longer times will give a better calibration.

Calibration failure alarm will issue a WARNING if the analyzer has to change calibration by more than the Cal failure error, if warning alarms are enabled.

Calibration time out sets how long the analyzer will wait for the signal to stabilize before issuing a WARNING. You can zero or span the ranges all at once or not.

HOME

 

ESCAPE

 

INFO

 

 

Menu: 109 INIT

Re-initialize the analyzer

Are you sure?

INIT will erase ALL the configuration data, including manufacturing data, serial numbers etc.

If you are sure, press INIT again.

 

 

HOME

ESCAPE

INIT

INFO

Menu: 110 SWDIAGI1

Software Diagnostics

Shows the first detected software error since the variable on the bottom line was reset.

Please report any errors to your service representative. They may mean nothing.

The analyzer has a lot of error recovery code.

Errors may therefore correct themselves.

HOME

 

ESCAPE

 

INFO

 

 