Model NGA2000 PMD

Instruction Manual

760007-A July 2003

Menu: 54 RN4FACTORS

 

 

 

Range 4 Factors

 

 

 

Zero offset:

 

 

 

 

 

 

 

000.0

Span factor:

 

 

 

 

 

 

 

000.0

Full scale range at calibration:

 

 

000.0

Measurement range number:

 

 

000.0

Hardware zero offset:

 

 

 

 

 

 

000.0

Raw measurement signal:

 

 

000.0

 

 

 

 

 

 

 

 

 

 

HOME

 

STORE

 

FIRST

 

 

HISTORY

 

INFO

Menu: 55 RFACTORSl

Range Factors

Shows the calibration factors for this range. Modify the zero factor for zero calibration, and the span factor for spanning this range. The take effect as soon as you press the enter key.

With zero gas, the raw reading should be the same as the zero offset.

Then do a complete recalibration.

HOME

 

ESCAPE

 

INFO

 

 

 

 

 

 

 

Menu: 56 AMHELPINDEX

Analyzer Module Help

Paramagnetic Oxygen detector

This device uses the paramagnetic nature of oxygen as a measurement technique. Other gases are usually diamagnetic and the detector responds only weakly to them. Nitrogen oxides are the only

common exceptions.

The analyzer is sensitive to vibration and should be installed with care.

HOME

 

ESCAPE

 

INFO

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendix A. Menu Displays