Model NGA2000 PMD
Instruction Manual
Menu: 54 RN4FACTORS
|
|
| Range 4 Factors |
|
|
| |||
Zero offset: |
|
|
|
|
|
|
| 000.0 | |
Span factor: |
|
|
|
|
|
|
| 000.0 | |
Full scale range at calibration: |
|
| 000.0 | ||||||
Measurement range number: |
|
| 000.0 | ||||||
Hardware zero offset: |
|
|
|
|
|
| 000.0 | ||
Raw measurement signal: |
|
| 000.0 | ||||||
|
|
|
|
|
|
|
|
|
|
HOME |
| STORE |
| FIRST |
|
| HISTORY |
| INFO |
Menu: 55 RFACTORSl
Range Factors
Shows the calibration factors for this range. Modify the zero factor for zero calibration, and the span factor for spanning this range. The take effect as soon as you press the enter key.
With zero gas, the raw reading should be the same as the zero offset.
Then do a complete recalibration.
HOME |
| ESCAPE |
| INFO |
|
| |||
|
|
|
|
|
Menu: 56 AMHELPINDEX
Analyzer Module Help
Paramagnetic Oxygen detector
This device uses the paramagnetic nature of oxygen as a measurement technique. Other gases are usually diamagnetic and the detector responds only weakly to them. Nitrogen oxides are the only
common exceptions.
The analyzer is sensitive to vibration and should be installed with care.
HOME |
| ESCAPE |
| INFO |
|
|
Rosemount Analytical Inc. A Division of Emerson Process Management | Appendix A. Menu Displays |