Model NGA2000 PMD

Instruction Manual

760007-A July 2003

Menu: 66 MPARMSI1

Current measurement parameters help

Shows the main measurement parameters. These can be controlled in the various set up menus.

HOME

 

ESCAPE

 

INFO

 

 

 

 

 

 

 

Menu: 67 BAROM_PARMS

 

 

Barometric pressure parameters

 

Pressure transducer:

000.0

Barometric pressure compensation:

000.0

Measured pressure:

000.0

Transducer offset:

000.0

Transducer slope:

000.0

Transducer PGA gain:

000.0

 

 

 

 

 

HOME

 

ESCAPE

 

INFO

Menu: 68 BAROM_PARMI1

Barometric pressure parameters

Shows whether the optional pressure transducer is installed or not. Also this screen allows you to set the transducer. Use the offset at atmospheric pressure, and pull a vacuum of about 25cm H2O on the transducer, and use the slope to make it read correctly.

Choose whether to enable barometric compensation, and whether to allow the analyzer to emit the pressure value. Only one analyzer at a time in a

HOME

 

ESCAPE

 

INFO

 

 

 

 

 

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendix A. Menu Displays

7-23