Model NGA2000 PMD
Instruction Manual
Menu: 66 MPARMSI1
Current measurement parameters help
Shows the main measurement parameters. These can be controlled in the various set up menus.
HOME |
| ESCAPE |
| INFO |
|
| |||
|
|
|
|
|
Menu: 67 BAROM_PARMS
|
| Barometric pressure parameters |
| |
Pressure transducer: | 000.0 | |||
Barometric pressure compensation: | 000.0 | |||
Measured pressure: | 000.0 | |||
Transducer offset: | 000.0 | |||
Transducer slope: | 000.0 | |||
Transducer PGA gain: | 000.0 | |||
|
|
|
|
|
HOME |
| ESCAPE |
| INFO |
Menu: 68 BAROM_PARMI1
Barometric pressure parameters
Shows whether the optional pressure transducer is installed or not. Also this screen allows you to set the transducer. Use the offset at atmospheric pressure, and pull a vacuum of about 25cm H2O on the transducer, and use the slope to make it read correctly.
Choose whether to enable barometric compensation, and whether to allow the analyzer to emit the pressure value. Only one analyzer at a time in a
HOME |
| ESCAPE |
| INFO |
|
| |||
|
|
|
|
|